User Contributed MET/CAL Procedure ============================================================================= INSTRUMENT: General Radio 1864:Megomhmeter DATE: 29-Aug-97 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 35 NUMBER OF LINES: 327 CONFIGURATION: Fluke 5700A CONFIGURATION: Datron 34401A STANDARD: DTB Megohm Resistance Box ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 STD DTB Megohm Resistance Box 1.004 HEAD * INITIAL CONDITIONS * 1.005 DISP Plug in the UUT and turn on the power. 1.005 DISP Allow the UUT to warm-up for « hour. 1.006 DISP [32] Connect the UUT as follows: 1.006 DISP + UNKNOWN ............................. 34401A HI 1.006 DISP - UNKNOWN ............................. 34401A LO 1.007 DISP [32] Set the UUT as follows: 1.007 DISP Test Voltage .......................... 1-0-V 1.007 DISP Discharge/Charge/Measure .............. Charge 1.012 HEAD {* Voltage Accuracy *} 1.013 JMP 2.002 1.014 EVAL 34401 10.000V 2.00% 2W 3.001 DISP [32] Set the UUT as follows: 3.001 DISP Test Voltage .......................... 2-0-V 34401 20.000V 2.00% 2W 4.001 DISP [32]Set the UUT as follows: 4.001 DISP Test Voltage .......................... 3-0-V 34401 30.000V 2.00% 2W 5.001 DISP [32]Set the UUT as follows: 5.001 DISP Test Voltage .......................... 4-0-V 34401 40.000V 2.00% 2W 6.001 DISP [32]Set the UUT as follows: 6.001 DISP Test Voltage .......................... 5-0-V 34401 50.000V 2.00% 2W 7.001 DISP [32]Set the UUT as follows: 7.001 DISP Test Voltage .......................... 6-0-V 34401 60.000V 2.00% 2W 8.001 DISP [32]Set the UUT as follows: 8.001 DISP Test Voltage .......................... 7-0-V 34401 70.000V 2.00% 2W 9.001 DISP [32]Set the UUT as follows: 9.001 DISP Test Voltage .......................... 8-0-V 34401 80.000V 2.00% 2W 10.001 DISP [32]Set the UUT as follows: 10.001 DISP Test Voltage .......................... 9-0-V 34401 90.000V 2.00% 2W 11.001 DISP [32]Set the UUT as follows: 11.001 DISP Test Voltage .......................... 10-0-V 34401 100.00V 2.00% 2W 13.001 DISP [32]Set the UUT as follows: 13.001 DISP Test Voltage .......................... 1-1-V 34401 11.000V 2.00% 2W 14.001 DISP [32]Set the UUT as follows: 14.001 DISP Test Voltage .......................... 1-2-V 34401 12.000V 2.00% 2W 15.001 DISP [32]Set the UUT as follows: 15.001 DISP Test Voltage .......................... 1-3-V 34401 13.000V 2.00% 2W 16.001 DISP [32]Set the UUT as follows: 16.001 DISP Test Voltage .......................... 1-4-V 34401 14.000V 2.00% 2W 17.001 DISP [32]Set the UUT as follows: 17.001 DISP Test Voltage .......................... 1-5-V 34401 15.000V 2.00% 2W 18.001 DISP [32]Set the UUT as follows: 18.001 DISP Test Voltage .......................... 1-6-V 34401 16.000V 2.00% 2W 19.001 DISP [32]Set the UUT as follows: 19.001 DISP Test Voltage .......................... 1-7-V 34401 17.000V 2.00% 2W 20.001 DISP [32]Set the UUT as follows: 20.001 DISP Test Voltage .......................... 1-8-V 34401 18.000V 2.00% 2W 21.001 DISP [32]Set the UUT as follows: 21.001 DISP Test Voltage .......................... 1-9-V 34401 19.000V 2.00% 2W 23.001 DISP [32]Set the UUT as follows: 23.001 DISP Test Voltage .......................... 10-0-0V 23.002 IEEE [@34401] DCV AUTO 23.003 ACC 2000 1000.00V 10P% +0.2P/ 23.004 IEEE [@34401] [D1000]RDG?[I] 23.005 MATH MEM1 = 1000 23.006 MEME 23.007 MEMC V 2.00% #! Test Tol 20, Sys Tol 0.0104, TUR 1923.077 (>= 4.00). 24.001 DISP [32]Set the UUT as follows: 24.001 DISP Discharge/Charge/Measure .............. Discharge 24.002 HEAD {} 24.003 HEAD {É*********»} 24.004 HEAD {º Resistance Tests º} 24.005 HEAD {È*********¼} 24.006 HEAD * Resistance Tests * 24.007 JMP 25.001 24.008 EVAL 25.001 DISP Disconnect the UUT from the Datron 34401A DMM. 25.001 DISP On the 5700 disconnect the GUARD to GND link. 25.002 DISP [32]Set the UUT as follows: 25.002 DISP Test Voltage .......................... 1-0-V 25.002 DISP Discharge/Charge/Measure .............. Measure 25.002 DISP Range Switch .......................... 100k 25.003 DISP With the UUT UNKNOWN terminals open, adjust the 25.003 DISP SET ì dial for an reading of ì on the UUT meter. 25.004 DISP [32]Set the UUT as follows: 25.004 DISP Discharge/Charge/Measure .............. Discharge 25.005 DISP Connect the UUT ñ UNKNOWN terminals to the Fluke 5700 25.005 DISP OUTPUT V/ê/A HI/LO terminals. 25.006 DISP [32]Set the UUT as follows: 25.006 DISP Discharge/Charge/Measure .............. Measure 25.007 HEAD {Test Voltage = 10} 25.008 JMP 26.001 25.009 EVAL 26.001 5700 100.000kZ S 2W 26.002 MEMI Enter the UUT meter reading. 26.003 MEM* 100 26.004 MEME 26.005 MEMC kZ 4.00% #! Test Tol 4000, Sys Tol 1.3, TUR 3076.923 (>= 4.00). 27.001 DISP [32]Set the UUT as follows: 27.001 DISP Range Switch .......................... 1M 27.002 5700 1.000MZ S 2W 27.003 MEMI Enter the UUT meter reading. 27.004 MEME 27.005 MEMC MZ 4.00% #! Test Tol 40000, Sys Tol 18, TUR 2222.222 (>= 4.00). 28.001 DISP [32]Set the UUT as follows: 28.001 DISP Range Switch .......................... 10M 28.002 5700 10.00MZ S 2W 28.003 MEMI Enter the UUT meter reading. 28.004 MEM* 10 28.005 MEME 28.006 MEMC MZ 4.00% #! Test Tol 4e+005, Sys Tol 370, TUR 1081.081 (>= 4.00). 29.001 MEME 29.002 DISP [32]Set the UUT as follows: 29.002 DISP Range Switch .......................... 100M 29.003 5700 100.0MZ S 2W 29.004 MEMI Enter the UUT meter reading. 29.005 MEM* 100 29.006 MEME 29.007 MEMC MZ 4.00% #! Test Tol 4e+006, Sys Tol 12000, TUR 333.333 (>= 4.00). 30.001 HEAD {} 30.002 HEAD {Test Voltage = 100} 30.003 JMP 31.001 30.004 EVAL 31.001 DISP [32]Set the UUT as follows: 31.001 DISP Discharge/Charge/Measure .............. Discharge 31.002 DISP Disconnect the UUT from the Fluke and connect it 31.002 DISP to the 1Gê resistor on the DTB MEGOHM resistance box. 31.003 DISP [32]Set the UUT as follows: 31.003 DISP Test Voltage .......................... 1-0-0V 31.003 DISP Discharge/Charge/Measure .............. Charge 31.003 DISP Range Switch .......................... 1G 31.004 ACC 1.0000Gê 2% 31.005 MEMI Enter the calibrated value for the 1G resistor. 31.006 MEME 31.007 DISP [32]Set the UUT as follows: 31.007 DISP Discharge/Charge/Measure .............. Measure 31.008 MEMI Enter the UUT meter reading. 31.009 MEME 31.010 MEMC Gê 4.0000% #! WARNING: Test Tol 0.04, Sys Tol 0.02, TUR 2.000 (< 4.00). 32.001 DISP [32]Set the UUT as follows: 32.001 DISP Discharge/Charge/Measure .............. Discharge 32.001 DISP Range Switch .......................... 10G 32.002 DISP Connect the UUT to 10Gê resistor 32.003 DISP [32]Set the UUT as follows: 32.003 DISP Discharge/Charge/Measure .............. Charge 32.004 ACC 10.000Gê 2% 32.005 MEMI Enter the calibrated value for the 10G resistor. 32.006 MEME 32.007 DISP [32]Set the UUT as follows: 32.007 DISP Discharge/Charge/Measure .............. Measure 32.008 MEMI Enter the UUT meter reading. 32.009 MEM* 10 32.010 MEME 32.011 MEMC Gê 4.0000% #! WARNING: Test Tol 0.4, Sys Tol 0.2, TUR 2.000 (< 4.00). 33.001 HEAD {} 33.002 HEAD {Test Voltage = 500} 33.003 JMP 34.001 33.004 EVAL 34.001 DISP [32]Set the UUT as follows: 34.001 DISP Discharge/Charge/Measure .............. Discharge 34.001 DISP Test Voltage .......................... 5-0-0V 34.001 DISP Range Switch .......................... 100G 34.002 DISP Connect the UUT to 100Gê resistor 34.003 DISP [32]Set the UUT as follows: 34.003 DISP Discharge/Charge/Measure .............. Charge 34.004 ACC 100.000Gê 2% 34.005 MEMI Enter the calibrated value for the 100G resistor. 34.006 MEME 34.007 DISP [32]Set the UUT as follows: 34.007 DISP Discharge/Charge/Measure .............. Measure 34.008 MEMI Enter the UUT meter reading. 34.009 MEM* 100 34.010 MEME 34.011 MEMC Gê 4.0000% #! WARNING: Test Tol 4, Sys Tol 2, TUR 2.000 (< 4.00). 35.001 DISP [32]Set the UUT as follows: 35.001 DISP Discharge/Charge/Measure .............. Discharge 35.002 HEAD {} 35.003 DISP {É****************Í»} 35.003 DISP {º THIS COMPLETES THE VERIFICATION º} 35.003 DISP {È****************ͼ} 35.004 END #! T.U.R.s less than 4.00: 3 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.